A Standard System Protocol
R-88
CJ-series Serial Communications Units Operation Manual for NJ-series CPU Unit(W494)
Additional Information
•
Use sequences No. 446 (High calibration set) and No. 447 (Low calibration set) to calibrate the
Laser Micrometer. If this sequence (No. 404) is used, both the high and low calibration gauges
must be set, and the high calibration and low calibration gauges cannot be exchanged.
•
The HC gauge dimension and LC gauge dimension must be set with 3 digits for the integer
portion and 4 digits for the decimal portion.
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
This sequence releases the calibration of the Laser Micrometer.
Send Data Word Allocation (3rd Operand of PMCR(260))
None.
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
Additional Information
In this sequence, you cannot separately release both high and low calibration.
This sequence switches the program number to a specified number.
Send Data Word Allocation (3rd Operand of PMCR(260))
Receive Data Word Allocation (4th Operand of PMCR(260))
None.
Additional Information
Retry processing is not performed for this sequence.
Calibration Release (Sequence No. 405 (Hex 0195))
Program Number Set (3000-series) (Sequence No. 406 (Hex 0196))
Offset
Contents (data format)
Data
+0
Number of send data words (4 digits Hex)
0002 (fixed)
+1
Program number (1 digit BCD)
0 to 9
First word of send
data
Number of send data words
+0
+1
(Undefined)
Program number (1 digit BCD)
Содержание CJ1W-SCU22
Страница 192: ...4 Using Protocol Macros 4 56 CJ series Serial Communications Units Operation Manual for NJ series CPU Unit W494...
Страница 250: ...6 No protocol Mode 6 22 CJ series Serial Communications Units Operation Manual for NJ series CPU Unit W494...
Страница 460: ...A Standard System Protocol R 126 CJ series Serial Communications Units Operation Manual for NJ series CPU Unit W494...
Страница 473: ...Index 1 CJ series Serial Communications Units Operation Manual for NJ series CPU Unit W494 I Index...