:VSOurce:RANGe <n>
This command is used to set the V-source range for Manual V-source Ohms.
Usage
[:SENSe[1]:RESistance:MANual:VSOurce:RANGe <n>
:RANGe?
:RANGe? DEFault
:RANGe? MINimum
:RANGe? MAXimum
<n>
0 to 1000:
Specify output in volts
DEFault:
100V range
MINimum:
100V range
MAXimum:
1000V range
Details
This command is used to select the optimum V-source range for Auto V-source Ohms. Specifying the
output voltage value automatically selects the most sensitive range. For example, specifying a value
less than or equal to 100 selects the 100 V range. Specifying a value greater than 100 selects the
1000 V range.
You cannot select the 100 V range if the programmed amplitude is greater than 100 V; a Settings
Conflict error results.