Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-44
6517B-901-01 Rev. C / August 2015
:CLIMits:FAIL?
This command reads the composite result of limit tests.
Usage
:CALCulate3:CLIMits:FAIL?
Details
This query command is used to obtain the composite result of the LIMIT 1 and LIMIT 2 tests. The
composite result is the logical ORed summary of LIMIT 1 and LIMIT 2:
1
= One or both tests have failed
0
= Both tests have passed
In the event of a failure (1), you can determine which test(s) has failed by using the fail command of
each limit (
:LIMit1:FAIL?
and
:LIMit2:FAIL?
).
This command (like other fail commands) does not clear the fail indication for LIMIT 1 or LIMIT 2.
Also see
None
BSTRobe:STATe <b>
This command controls the binning strobe.
Usage
:CALCulate3:BSTRobe:STATe <b>
<b>
1 or ON:
Enable binning strobe
0 or OFF:
Disable binning strobe
Details
This command is used to enable or disable the strobe for the binning operation. When enabled, digital
output line #4 is used for the binning strobe pulse. After the appropriate digital pattern is placed on
the other three lines of the output port, the >10 µsec strobe pulse is asserted on line #4. The strobe is
used to "inform" your external binning circuit that the output port (lines 1, 2 and 3) is ready to be read.
Also see
None