Model 6517B Electrometer Reference Manual
Section 7: Triggering
6517B-901-01 Rev. C / August 2015
7-19
Asynchronous trigger link example
In a typical test system, you may want to close a channel and then measure the DUT connected to
the channel with a meter. Such a test system is shown in the following figure, which uses a Model
6517B Electrometer to measure ten DUTs switched by a 10-channel multiplexer card in a Model
7001/7002 Switch System.
Model 6517B:
Idle State:
Bench reset = :INIT:CONT ON*
Arm layer:
Arm source = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
Scan layer:
Scan source = Immediate*
Scan count = Infinite*
Scan trigger control = Acceptor*
Measure layer:
Measure source = TrigLink
Trigger link mode = Asynchronous
Input line = #3
Output line = #4
Measure count = 20
Measure trigger control = Acceptor*
* Indicates that the setting is the BENCH RESET (and factory) default condition.
Figure 82: DUT test system