![background image](http://html2.mh-extra.com/html/keithley/6517b/6517b_reference-manual_4068717058.webp)
Model 6517B Electrometer Reference Manual
Section 3: Connections
6517B-901-01 Rev. C / August 2015
3-5
V-SOURCE LO connected to ammeter input LO through the METER-CONNECT option of
CONFIGURE V-SOURCE Menu.
The equivalent circuit for this configuration is shown in the following figure.
Figure 15: Force voltage measure current - equivalent circuit
Voltage source output connections
The voltage source output is accessed at the rear panel V SOURCE OUT HI and LO binding posts as
shown in the following figure.
Figure 16: V-source output basic connections
Using these terminals places the independent V-source in series with the external circuit (RL), as
shown in the following figure.
Figure 17: V-source output - equivalent circuit