Model 6517B Electrometer Reference Manual
Section 12: Common commands
6517B-901-01 Rev. C / August 2015
12-11
*SAV (save the current setup in memory)
This command saves the Model 6517B present setup configuration in memory for later recall.
Usage
*SAV <NRf>
<NRf>
Specify memory location:
0
to
9
Details
The
*SAV
command is used to save the current instrument setup configuration in memory for later
recall. Any control affected by
*RST
can be saved by the
*SAV
command. The
*RCL
command is
used to restore the instrument to a saved setup configuration.
Also see
*RCL (recall)
(on page 12-10)
*SRE <NRf> (service request enable), *SRE? (service request
enable query)
These commands program and read the service request enable register.
Usage
*SRE <NRf>
*SRE?
<NRf>
0
: Clears enable register
1
: Set MSB bit (Bit 0)
4
: Set EAV bit (Bit 2)
8
: Set QSB bit (Bit 3)
16
: Set MAV bit (Bit 4)
32
: Set ESB bit (Bit 5)
128
: Set OSB bit (Bit 7)
255
: Sets all bits
Details
The
*SRE
command is used to program the service request enable register. This command is sent
with the decimal equivalent of the binary value that determines the desired state (0 or 1) of each bit in
the register. This register is cleared on power-up.
This enable register is used along with the status byte register to generate service requests (SRQ).
With a bit in the service request enable register set, an SRQ occurs when the corresponding bit in the
status byte register is set by an appropriate event. For details on register structure, refer to
Status
structure
(on page 13-1).