![background image](http://html2.mh-extra.com/html/keithley/6517b/6517b_reference-manual_4068717450.webp)
Model 6517B Electrometer Reference Manual
Section 15: Calibration procedure
6517B-901-01 Rev. C / August 2015
15-27
Error Code
Description
+321
Buffer sizing error; set to MAX
+322
Buffer sizing error; set to MIN
+350
2 V offset out of specification
+351
2 V positive gain out of specification
+352
2 V negative gain out of specification
+353
20 V offset out of specification
+354
20 V positive gain out of specification
+355
20 V negative gain out of specification
+356
200 V offset out of specification
+357
200 V positive gain out of specification
+358
200 V negative gain out of specification
+359
20 pA offset out of specification
+360
20 pA positive gain out of specification
+361
20 pA negative gain out of specification
+362
200 pA offset out of specification
+363
200 pA positive gain out of specification
+364
200 pA negative gain out of specification
+365
2 nA offset out of specification
+366
2 nA positive gain out of specification
+367
2 nA negative gain out of specification
+368
20 nA offset out of specification
+369
20 nA positive gain out of specification
+370
20 nA negative gain out of specification
+371
200 nA offset out of specification
+372
200 nA positive gain out of specification
+373
200 nA negative gain out of specification
+374
2 mA offset out of specification
Содержание 6517B
Страница 136: ...Section 6 Test sequences Model 6517B Electrometer Reference Manual 6 8 6517B 901 01 Rev C August 2015...
Страница 180: ...Section 9 Filters and math Model 6517B Electrometer Reference Manual 9 4 6517B 901 01 Rev C August 2015...
Страница 492: ...In this appendix Information E 2 Appendix E IEEE 488 conformance...