Section 15: Calibration procedure
Model 6517B Electrometer Reference Manual
15-28
6517B-901-01 Rev. C / August 2015
Error Code
Description
+375
2 mA positive gain out of specification
+376
2 mA negative gain out of specification
+377
20 mA offset out of specification
+378
20 mA positive gain out of specification
+379
20 mA negative gain out of specification
+380
200 mA offset out of specification
+381
200 mA positive gain out of specification
+382
200 mA negative gain out of specification
+383
2 mA offset out of specification
+384
2 mA positive gain out of specification
+385
2 mA negative gain out of specification
+386
20 mA offset out of specification
+387
20 mA positive gain out of specification
+388
20 mA negative gain out of specification
+389
2 nC positive gain out of specification
+390
2 nC negative gain out of specification
+391
20 nC positive gain out of specification
+392
20 nC negative gain out of specification
+393
200 nC positive gain out of specification
+394
200 nC negative gain out of specification
+395
2 mC positive gain out of specification
+396
2 mC negative gain out of specification
+397
Charge zero excessive
+399
Temp offset out of specification
+400
Temp gain out of specification
+410
Humidity offset out of specification
+412
Humidity gain out of specification