Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-134
6517B-901-01 Rev. C / August 2015
:LLOCkout <b>
This command enables or disables local lockout.
Usage
:SYSTem:LLOCkout <b>
:LLOCkout?
<b>
0
or
OFF
: Disable local lockout
1
or
ON
: Enable local lockout
Details
If enabled, the front-panel keys are locked out (non-operational) when the instrument is in remote. If
disabled, the front panel keys are operational in remote. The instrument must be in remote in order to
use this command.
Taking the instrument out of remote restores front-panel key operation, but does not change the
status of the
:LLOCkout
command.
Note that this command can only be sent over the RS-232 interface.
Also see
:LOCal
(on page 14-133)
:REMote
(on page 14-133)
Basic trigger commands
The following commands are used to configure the Model 6517B for basic triggering. See
Triggering
(on page 7-1) for details on basic triggering.
[:EXECute]
This command exits from advanced triggering and selects basic triggering.
Usage
:SYSTem:MACRo:TRIGger:MODE
Details
When this action command is sent, the instrument exits from advanced triggering and select basic
triggering. Sending this command while already in basic triggering has no effect on the instrument.
Also see
None