Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-126
6517B-901-01 Rev. C / August 2015
The
:KEY
command allows you to select the NEXT displays over the bus. Sending
:SYSTem:KEY
16
over the bus is same as a single press of the
NEXT
key. Sending
:SYSTem:KEY 8
is the same
as a single press of the
PREV
key. The NEXT display that is selected depends on how many times
the
NEXT
key is pressed using the
:KEY
command. The instrument can always be returned to the
normal display by pressing the
PREV
key a similar number of times, or by sending the
:DISPlay:CNDisplay
command.
For example, you can display the bar graph by pressing the
NEXT
key twice as follows:
:syst:key 16; key 16
To return to the normal display, press the
PREV
key twice as follows:
:syst:key 8; key 8
The queue for the
:KEY?
query command can only hold one key-press. When
:KEY?
is sent over the
bus, and the Model 6517B is addressed to talk, the key-press code number for the last key pressed
(either physically or with
:KEY
) is sent to the computer.
Figure 134: Key press codes
Also see
None
:CLEar
This command is used to clear the error queue.
Usage
:SYSTem:CLEar
Details
This action command is used to clear the error queue of messages.
Also see
None