Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-35
:REFerence <NRf>
This command is used to set the reference value for ratio, deviation, and percent deviation
Usage
:CALCulate[1]:REFerence <NRf>
:REFerence?
<NRf>
-9.999999e35 to +9.999999e35:
Specify reference value
Details
This command is used to specify the reference value for the ratio, deviation, and percent deviation
calculation.
Also see
None
:STATe <b>
This command is used to control the CALC 1 calculation.
Usage
:CALCulate[1]:STATe <b>
:STATe?
<b>
0
or
OFF:
Disable CALC 1 calculation
1
or
ON:
Enable CALC 1 calculation
Details
This command is used to enable or disable the CALC 1 calculation. When enabled, each instrument
reading reflects the selected calculation.
Also see
:FORMat <name>
(on page 14-37)
DATA commands
[:LATest]?
This command is used to read the
CALCulate[1]
result.
Usage
:CALCulate[1]:DATA[:LATest]?
Details
This query command is used to read the result of the
CALCulate[1]
calculation. If
CALCulate[1]
is disabled or
NONE
is selected, the "raw" reading is read instead.
Also see
CALCulate[1]
(on page 14-32)