Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-36
6517B-901-01 Rev. C / August 2015
:FRESh?
This command is used to recalculate
CALCulate[1]
.
Usage
:CALCulate[1]:DATA:FRESh?
Details
This query command is used to return a new (fresh) math result of
CALCulate[1]
. This command
does not request the same result twice. If a new reading is triggered, this command waits until the
math result is available. When
CALCulate[1]
is disabled, there is no difference between
SENS:DATA:FRES?
and
CALC:DATA:FRES?.
Also see
CALCulate[1]
(on page 14-32)
:IMMediate
This command is used to recalculate CALC1.
Usage
:CALCulate[1]:IMMediate
:IMMediate?
Details
When you change the math calculation, the next instrument reading reflects the new calculation
(assuming
CALC 1
math is enabled). If the instrument is not in a continuous measurement mode (for
example, waiting for a trigger), the displayed reading does not update to reflect the new calculation
until the next reading conversion occurs.
This action command allows you to re-process the current input data to reflect a new calculation. For
example, assume the instrument is in a non-continuous measurement mode and requires a manual
trigger to cause the next reading conversion. Changing the calculation does not affect the currently
displayed reading. However, sending the
:IMMediate
command reprocesses that displayed reading
to reflect the new calculation. Note that sending the
:IMMediate
command does not initiate a
reading conversion.
An alternate way to perform the calculation and read the result is by using the query form of the
command (
:IMMediate?
). When this command is sent, the calculation is performed and the result is
queried.
Also see
None
CALCulate2
This calculate subsystem is used to configure and control
CALC 2
operations on readings stored in the buffer.
Содержание 6517B
Страница 136: ...Section 6 Test sequences Model 6517B Electrometer Reference Manual 6 8 6517B 901 01 Rev C August 2015...
Страница 180: ...Section 9 Filters and math Model 6517B Electrometer Reference Manual 9 4 6517B 901 01 Rev C August 2015...