Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-78
6517B-901-01 Rev. C / August 2015
:NTOLerance <n>
This command is used to specify the noise tolerance for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage:ADVanced:NTOLerance <n>
[:SENSe[1]]:CURRent[:DC]:AVERage:ADVanced:NTOLerance <n>
[:SENSe[1]:RESistance:AVERage:ADVanced:NTOLerance <n>
[:SENSe[1]]:CHARge:AVERage:ADVanced:NTOLerance <n>
:NTOLerance?
:NTOLerance? DEFault
:NTOLerance? MINimum
:NTOLerance? MAXimum
<n>
1 to 100:
Specify noise tolerance in percent (+/-)
DEFault:
+/-1 %
MINimum:
0 %
MAXimum:
+/-100 %
Details
Each measurement function can utilize the advanced filter which adds a noise tolerance window to
the averaging algorithm. These commands are used to set the noise tolerance level of the advanced
filter for the specified function. The
:ADVanced[:STATe]
commands are used to enable or disable
the advanced filter.
The noise tolerance level is expressed as a percent of the last averaged reading. In general, if the
noise is within this window, the reading is simply based on the normal averaging algorithm. If a
reading is outside this window, then the advanced averaging algorithm is used.
Also see
None