Section 16: Verification procedure
Model 6517B Electrometer Reference Manual
16-12
6517B-901-01 Rev. C / August 2015
2. Turn on the power to the Model 6517A and allow it to warm up for at least two hours before
making measurements.
3. Enable the Model 6517B humidity measurements by turning on the humidity data stamp.
a. Press
MENU
, use the right arrow key to select
GENERAL
on the second page of the lower
display, and then press
ENTER
.
b. Use the right arrow key to select
A/D-CONTROLS
, and then press
ENTER
.
c. Select
DATA-STAMP
, and then press
ENTER
.
d. Use the left or right arrow keys to select
HUMIDITY
, and then use the range up or range down key to
select
ON
.
e. Press
ENTER
, and then press
EXIT
three times to return to the measurement display.
f. Press
PREV
to view the relative humidity (RH) reading.
4. Set the calibrator output to 0.0 V.
5. Allow the reading to settle, then verify that the Model 6517B humidity reading is within the limits in
the following table.
6. Repeat step 5 for each of the voltage setting list in the table.
Limits for humidity verification
Applied
voltage
Humidity reading limits
(1 year, 18° to 28°C)
0.0 V
0 % to 1 %
0.25 V
24 % to 26 %
0.5 V
49 % to 51 %
0.75 V
74 % to 76 %
1.0 V
99 % to 100 %
Ohms verification
Ohms function accuracy is verified by connecting accurate resistance standards to the Model 6517B
and then verifying that the readings on the display fall within the required ranges. The following
paragraphs discuss the required resistance standards, how to compute voltage source values, and
provide detailed procedures for verifying accuracy of the ohms function.
Ohms specifications are derived from amps and voltage source specifications. Thus, it is not
necessary to verify ohms separately. However, the following procedure is provided if you want to
perform an Artifact Standard Verification.
Required standard resistors
Resistance standards required include:
Resistance calibrator (2 M
Ω
to 200 M
Ω
ranges)
Model 5156 Electrometer Calibration Standard (2 G
Ω
to 200 G
Ω
ranges)
Characterized resistors (2 T
Ω
to 200 T
Ω
ranges)
See the
Equipment needed for verification
(on page 16-2) topic for detailed information on the
recommended standard resistors.