Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-3
:CONFigure:<function>
This command configures the instrument for subsequent measurements on the specified function.
Usage
:CONFigure:<function>
:CONFigure?
<function>
The function:
CURRent[:DC]
: Amps function
RESistance
: Ohms function
CHARge
: Coulombs function
Details
This command is used to configure the instrument for subsequent measurements on the specified
function. Basically, this command places the instrument in a one-shot measurement mode. The
:READ?
command can then be used to trigger a measurement and acquire a reading.
When this command is sent, the Model 6517B is configured as follows:
The function specified by this command is selected.
All controls related to the selected function are defaulted to the
*RST
values.
Continuous initiation is disabled (
:INITiate:CONTinuous OFF
).
The control sources in all three layers of the trigger model are set to Immediate.
The count values in all three layers of the trigger model are set to one (1).
The delay in all three layers of the trigger model are set to zero (0).
The Model 6517B is placed in the idle state.
All math calculations are disabled.
Buffer operation is disabled; a storage operation currently in process is aborted.
Autozero controls are set to the
*RST
default values.
The acquisition method is set to normal.
All operations associated with switching cards (scanning) are disabled.
This command is automatically asserted when the
:MEASure?
command is sent.
Also see
READ?
(on page 14-4)