Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-19
Command Description
Default
parameter
SCPI
:TYPE <name>
Select filter type (
NONE, SCALar,
ADVanced
)
NONE
:TYPE?
Query filter type
:TCONtrol
<name>
Select filter type: (
MOVing
or
REPeat
)
MOVing
:TCONtrol?
Query filter type
:COUNt <n>
Specify filter count (1 to 100)
10
:COUNt?
Query filter count
:ADVanced
Path to configure and control advanced
filter
:NTOLerance
<n>
Specify noise tolerance level (0 to 100 %) 1
:NTOLerance?
Query noise tolerance level
[:STATe] <b>
Enable or disable filter
OFF
[:STATe]?
Query state of digital filter
:MEDian
Path to control the median filter:
[:STATe] <b>
Enable or disable median filter
ON
[:STATe]?
Query state of median filter
:RANK <NRf>
Specify rank for median filter
1
:RANK?
Query rank
:ADIScharge
Path to control auto discharge:
[:STATe] <b>
Enable or disable auto discharge
OFF
[:STATe]?
Query state of auto discharge
:LEVel <NRf>
Specify level (-2.2e6 to 2.2e6)
2e-6
:LEVel?
Query level
NOTES:
1. 16.67 ms for 60 Hz power line frequency and 20ms for 50 Hz line frequency.
2.
*RST
— OFF,
:SYSTem:PRESet
— ON.
3.
*RST
—
REPeat, :SYSTem:PRESet
— MOVing.
4.
*RST
— ON, :
SYSTem:PRESet
— OFF.
Содержание 6517B
Страница 136: ...Section 6 Test sequences Model 6517B Electrometer Reference Manual 6 8 6517B 901 01 Rev C August 2015...
Страница 180: ...Section 9 Filters and math Model 6517B Electrometer Reference Manual 9 4 6517B 901 01 Rev C August 2015...
Страница 492: ...In this appendix Information E 2 Appendix E IEEE 488 conformance...