Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-11
OUTPut command summary
Command Description
Default
Parameter
SCPI
:OUTPut1
[:STATe] <b>
Enable (operate) or disable (standby)
OFF
V-source output
[:STATe]?
Query state of V-source output
:TTL[1]
Path to set polarity of digital output line 1:
:LSENse <name>
Select polarity; active high (
AHIGh
)
AHIGh
or active low (
ALOW
)
:LSENse?
Query polarity of line 1
:TTL2
Path to set polarity of digital output line 2:
:LSENse <name>
Select polarity; active high (
AHIGh
)
AHIGh
or active low (
ALOW
)
:LSENse?
Query polarity of line 2
:TTL3
Path to set polarity of digital output line 3:
:LSENse <name>
Select polarity; active high (AHIGh)
AHIGh
or active low (ALOW)
:LSENse?
Query polarity of line 3
:TTL4
Path to set polarity of digital output line 4:
:LSENse <name>
Select polarity; active high (
AHIGh
)
AHIGh
or active low (
ALOW
)
:LSENse?
Query polarity of line 4
:ROUTe command summary
Command Description
Default
parameter
SCPI
:ROUTe
:CLOSe <list>
Path and command to close specified channel:
:STATe?
Query closed channel
:CLOSe? <list>
Query specified channels (
1
= closed,
0
= open)
:OPEN <list>
Open specified channel or all channels
:OPEN:ALL
Open all channels
:OPEN? <list>
Query specified channels (
1
= open,
0
= closed)
:SCAN
Command path to configure and control
scan operations:
Содержание 6517B
Страница 136: ...Section 6 Test sequences Model 6517B Electrometer Reference Manual 6 8 6517B 901 01 Rev C August 2015...
Страница 180: ...Section 9 Filters and math Model 6517B Electrometer Reference Manual 9 4 6517B 901 01 Rev C August 2015...
Страница 492: ...In this appendix Information E 2 Appendix E IEEE 488 conformance...