Section 7: Triggering
Model 6517B Electrometer Reference Manual
7-24
6517B-901-01 Rev. C / August 2015
The two instruments are configured as follows:
Model 6517B:
Idle state:
Bench reset = :INIT:CONT ON*
Arm layer:
Arm source = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
Scan layer:
Scan source = Immediate*
Scan count = Infinite*
Scan trigger control = Acceptor*
Measure layer:
Measure source = TrigLink
Trigger link mode = Semi-synchronous
Semi-sync line = #1*
Measure count = 10
Measure trigger control = Acceptor*
* Indicates that the setting is the BENCH RESET (and factory) default condition.
Model 7001 or 7002:
Idle state:
Reset = :INIT:CONT OFF*
Scan list = 1!1-1!10,
Arm layer:
Arm spacing = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
Scan layer:
Scan spacing = Immediate*
Number of scans = 1
Scan trigger control = Acceptor*
Channel layer:
Channel spacing = TrigLink
Trigger link mode = Semi-synchronous
Semi-sync line = #1
Number of channels = Use Scanlist length*
Channel trigger control = source*
* Indicates that the setting is the RESET (and factory) default condition.
To run the test and store the readings in the Model 6517B, press
STORE
on the electrometer, enter
the desired number of readings (ten), and press
ENTER
. The Model 6517B waits (with the asterisk
annunciator lit) for a trigger link trigger from the Model 7001/7002. Press
STEP
on the Model
7001/7002 to start the scan.