Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-37
:FORMat <name>
This command is used to specify the CALC 2 format.
Usage
:CALCulate2:FORMat <name>
:FORMat?
<name>
MEAN:
Mean value of readings in buffer
SDEV:
Standard deviation of readings in buffer
MAX:
Largest reading in buffer
MIN:
Lowest reading in buffer
PKPK:
Peak-to-peak value of readings in buffer
NONE:
No calculation
Details
Description This command is used to specify the format for the
CALC 2
math calculation. The
calculation operations for
CALC 2
use data stored in the buffer.
With
NONE
selected, no
CALC 2
calculation is performed. With one of the other formats selected and
CALC 2
enabled, the calculation is performed every time the
:IMMediate
or
:IMMediate?
command is executed.
PKPK = MAX - MIN
Where:
MAX
is the largest reading in the buffer, and MIN is the lowest reading in the buffer.
Also see
:IMMediate
(on page 14-38)
:STATe <b>
(on page 14-37)
:STATe <b>
This command is used to control
CALC 2
.
Usage
:CALCulate2:STATe <b>
:STATe?
<n>
0 or OFF:
Disable
CALC 2
calculation
1 or ON:
Enable
CALC 2
calculation
Details
This command is used to enable or disable the
CALC 2
calculation. When enabled, the selected
CALC 2
format is calculated when the
:IMMediate
or
:IMMediate?
command is executed.
Also see
:IMMediate