![background image](http://html2.mh-extra.com/html/keithley/6517b/6517b_reference-manual_4068717361.webp)
Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-110
6517B-901-01 Rev. C / August 2015
Measurement event
Positive transition effect on transition filter event register
Reading overflow
Sets B0 when reading exceeds range limits
Low limit 1
Sets B1 when reading is less than the low limit 1 setting
High limit 1
Sets B2 when reading is greater than the low limit 1 setting
Low limit 2
Sets B3 when reading is less than the low limit 2 setting
High limit 2
Sets B4 when reading is greater than the high limit 2 setting
Reading available
Sets B5 after a reading is taken and processed
Reading underflow
Sets B6 when an ohms reading is below the lower range limit
Buffer available
Sets B7 when the trace buffer has at least two readings
Buffer half-full
Sets B8 when the trace buffer becomes half-full
Buffer full
Sets B9 when the trace buffer becomes full
Sequence reading available
Sets B10 when an alternating polarity test sequence is being read
Buffer pretriggered
Sets B11 when the trace buffer pretrigger event occurs
Out of Limits
Sets B12 when a noise spike occurs
Fixture lid closed
Sets B13 when fixture lid is closed
V-source compliance
Sets B14 when V-source goes in compliance
Figure 128: Transition filter