Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-76
6517B-901-01 Rev. C / August 2015
:TYPE <name>
This command is used to select the filter type for volts, amps, ohms, and volts.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage:TYPE <name>
[:SENSe[1]]:CURRent[:DC]:AVERage:TYPE <name>
[:SENSe[1]:RESistance:AVERage:TYPE <name>
[:SENSe[1]]:CHARge:AVERage:TYPE <name>
:TYPE?
<name>
NONE:
No digital filter selected
SCALar:
Select scalar filter
ADVanced:
Select advanced filter
Details
Description This command is used to select the digital filter type. The
SCALar
filter is a simple
averaging type filter. The
ADVanced
filter is similar except that it incorporates a noise window to allow
faster response time to large signal step changes. The
:NTOLerance
command is used to set the
window for the
ADVanced
filter.
The number of readings that are averaged by the filter is set with the :
COUNt
command, and the
:
TCONtrol
command is used to select the filter mode (moving or repeating). The
[:STATe]
command is used to enable or disable the filter.
Setting
TYPE
to
NONE
with
MEDian:STATe
set to
OFF
sets
AVER:STATe
to
OFF
.
Also see
None
:TCONtrol <name>
This command is used to select the filter mode for volts, amps, ohms, and coulombs.
Usage
[:SENSe[1]]:VOLTage[:DC]:AVERage:TCONtrol <name>
[:SENSe[1]]:CURRent[:DC]:AVERage:TCONtrol <name>
[:SENSe[1]:RESistance:AVERage:TCONtrol <name>
[:SENSe[1]]:CHARge:AVERage:TCONtrol <name>
:TCONtrol?
<name>
REPeat:
Select repeating filter mode
MOVing:
lect moving filter mode
Details
These commands are used to select the filter mode (
REPeat
or
MOVing
) for the specified function
(see
Digital filters
(on page 9-2) for more information).
MOV
during scan or test sequence generates a "Settings Conflict Error."
TCON
is set to
REPeat
during
a scan or test sequence if filter is
ON
and the appropriate function is encountered.
Also see
None