Input capacitance
At very high resistance levels, the very large time constants created by even a minimal amount of
capacitance can slow down response time considerably. For example, measuring a source with an
internal resistance of 100 G
would result in an RC time constant of one second when measured
through a cable with a nominal capacitance of 10 pF. If 1 % accuracy is required, a single
measurement would require at least five seconds.
There are two ways to minimize this problem:
Keep the input cable as short as possible
Use guarding
There is a limit to how short the cable can be. Using guard can reduce these effects by up to a factor
of 1000 (refer to
Guarding
(on page 4-16)).
Guarding
Guarding should be used for high-impedance voltage measurements and for voltage measurements
that use long input cables.
To understand the concept of guarding, review the unguarded circuit shown in the figure below. E
S
and R
S
represent the resistance and voltage components of the source, and R
L
and C
L
represent the
leakage resistance and cable capacitance of the triaxial input cable. The equivalent circuit shows the
divider that is formed. If R
S
is large enough, the divider significantly attenuates the voltage seen at the
input of the Model 6517B (see
Cable leakage resistance
(on page 4-6)). Also, R
S
and the cable
capacitance (C
L
) could create a long RC time constant, resulting in a slow measurement response
(see
Input capacitance
(on page 4-7)).