Model 6517B Electrometer Reference Manual
Section 15: Calibration procedure
6517B-901-01 Rev. C / August 2015
15-21
Command Description
:CAL:PROT:AZERO20M
Calibrate the 20 mA range zero
:CAL:PROT:A20M <+20 mA value>
Calibrate the +20 mA range
:CAL:PROT:AN20M <-20 mA value>
Calibrate the -20 mA range
Voltage measurement calibration commands
This is a list of all Model 6517B voltage measurement calibration commands.
Voltage measurement calibration commands
Command Description
:CAL:PROT:VZERO2
Calibrate the 2 V range zero
:CAL:PROT:V2 <+2 V value>
Calibrate the +2 V range
:CAL:PROT:VN2 <-2 V value>
Calibrate the -2 V range
:CAL:PROT:VZERO20
Calibrate the 20 V range zero
:CAL:PROT:V20 <+20 V value>
Calibrate the +20 V range
:CAL:PROT:VN20 <-20 V value>
Calibrate the -20 V range
:CAL:PROT:VZERO200
Calibrate the 200 V range zero
:CAL:PROT:V200 <+200 V value>
Calibrate the +200 V range
:CAL:PROT:VN200 <-200 V value>
Calibrate the -200 V range