Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-26
6517B-901-01 Rev. C / August 2015
:TRIGger command summary
Command Description
Default
parameter
SCPI
:INITiate
Subsystem command path:
[:IMMediate]
Initiate one trigger cycle
:CONTinuous <b>
Enable or disable continuous initiation
(Note 1)
:CONTinuous?
Query continuous initiation
:POFLag <name>
INCLude
or
EXCLude
no pending flag
for initiate commands
INCLude
:POFLag?
Query pending operation flag
:ABORt
Reset trigger system
:ARM[:SEQuence[1]]
Path to configure arm layers:
[:LAYer[1]]
Path to program Arm Layer 1:
:IMMediate
Loop around control source
:COUNt <n>
Specify arm count (1 to 99999, or INF) 1
:COUNt?
Query arm count
:SOURce <name>
Select control source (
HOLD,
IMMediate
,
RTCLock
,
MANual
,
BUS
,
TLINk
, or
EXTernal
)
IMMediate
(Note 2)
:SOURce?
Query control source
:SIGNal
Loop around control source
:TCONfigure
Path to configure triggers:
:DIRection
<name>
Enable (
SOURce
) or disable
(
ACCeptor
)
ACCeptor
:DIRection?
Query direction
:ASYNchronous
Path to configure asynchronous
trigger link:
:ILINe
<NRf>
Select input line (1 to 6)
2
:ILINe?
Query input line
:OLINe
<NRf>
Select output line (1 to 6)
1
:OLINe?
Query output line
:RTCLock
Path to configure the
RTCLock
control source:
:DATE <yr>,
<mo>, <day>
Specify the date for the clock event
:DATE?
Query the date for the clock event
:TIME <hr>,
<min>, <sec>
Specify the time for the clock event
:TIME?
Query the time for the clock event
:LAYer2
Path to program Arm Layer 2:
:IMMediate
Loop around control source
:COUNt <n>
Specify scan count (1 to 99999 or INF) (Note 3)
:COUNt?
Query scan count
Содержание 6517B
Страница 136: ...Section 6 Test sequences Model 6517B Electrometer Reference Manual 6 8 6517B 901 01 Rev C August 2015...
Страница 180: ...Section 9 Filters and math Model 6517B Electrometer Reference Manual 9 4 6517B 901 01 Rev C August 2015...
Страница 492: ...In this appendix Information E 2 Appendix E IEEE 488 conformance...