![background image](http://html2.mh-extra.com/html/keithley/6517b/6517b_reference-manual_4068717289.webp)
Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-38
6517B-901-01 Rev. C / August 2015
:IMMediate
This command is used to perform
CALC 2
.
Usage
:CALCulate2:IMMediate
:IMMediate?
Details
The
:IMMediate
command is used to perform the selected
CALC 2
operation on the readings in the
buffer (assuming
CALC 2
is enabled). After performing the calculation, the result can be read by
using the
:CALCulate2:DATA?
query command.
An alternate way to perform the calculation and read the result is by using the query form of the
command (
:IMMediate?
). When this command is sent, the calculation is performed and the result is
queried.
Also see
None
DATA?
This command is used to read the
CALC 2
result.
Usage
:CALCulate2:DATA?
Details
This query command is used to read the result of the
CALC 2
operation. If
CALC 2
is disabled or
NONE
is selected, the "raw" reading is read instead.
Note that this command does not initiate a calculation. This command is used only to read the result
of the last calculation.
Also see
None
CALCulate3
This Calculate subsystem is used to configure and control the limit tests (LIMIT 1 test and LIMIT 2 test).