Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-81
[:STATe] <b>
This command is used to control the median volts, amps, ohms, and coulombs.
Usage
[:STATe] <b>
[:SENSe[1]]:VOLTage[:DC]:MEDian[:STATe] <b>
[:SENSe[1]]:CURRent[:DC]:MEDian[:STATe] <b>
[:SENSe[1]]:RESistance:MEDian[:STATe] <b>
[:SENSe[1]]:CHARge:MEDian[:STATe] <b>
[STATe]?
<b>
1 or ON:
Enable median filter
0 or OFF
: Disable median filter
Details
This command is used to control (on or off) the median filter. When enabled, the instrument determines and
displays the median reading for the specified number of readings. The
:RANK
command is used to specify the
number of readings used to determine the median.
Note that the median filter can be used with the average filter. When both of these filters are enabled, each
displayed reading reflects the median for the average readings.
Setting
STATe
to
OFF
with
AVER:TYPE
set to
NONE
sets
AVER:STAT
to
OFF
.
Also see
None