Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-138
6517B-901-01 Rev. C / August 2015
:POINts <n>
This command is used to specify the buffer size.
Usage
:TRACe:POINts <n>
:POINts?
<n>
<n> =: 1 to 50000:
Specify number or readings
DEFault:
100
MAXimum:
See
Data elements
(on page 8-1).
MINimum:
1
Details
This command is used to specify the size of the buffer. When the specified number of readings are
stored in the buffer, the buffer is considered to be full.
For compatibility with the 6517A, the
MAX
parameter results in buffer sizes that match the 6517A.
The Model 6517B allows up to 50,000 points to be used if specified.
Also see
None