Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-8
6517B-901-01 Rev. C / August 2015
:FORMat?
Query math format
(Note)
:STATe <b>
Enable or disable calculation
:STATe?
Query state of math function
:IMMediate
Recalculate raw input data in buffer
:DATA?
Read math result of
CALC 2
:CALCulate3
Subsystem to control
CALC 3
(limit tests):
:LIMit[1]
Path to control
LIMIT 1
test:
:UPPer
Path to configure upper limit:
[:DATA] <n>
Set upper limit (-9.999999e35 to
1
+9.999999e35)
[:DATA]?
Query upper limit
0
:SOURce <NRf>
Specify digital output (0 to 15)
:SOURce?
Query source
:LOWer
Path to configure lower limit:
[:DATA] <n>
Set lower limit (-9.999999e35 to
-1
+9.999999e35)
[:DATA]?
Query lower limit
:SOURce <NRf>
Specify digital output (0 to 15)
0
:SOURce?
Query source
:STATe <b>
Enable or disable limit test
OFF
:STATe?
Query state of limit test
:FAIL?
Query test result (1 = pass, 0 = fail)
:CLEAR
Path to clear failed test:
[:IMMediate]
Clear failed test indication
:AUTO <b>
Enable or disable auto clear
ON
:AUTO?
Query auto clear
:LIMit2
Path to control LIMIT 2 test:
:UPPer
Path to configure upper limit:
[:DATA] <n>
Set upper limit (-9.999999e35 to
1
+9.999999e35)
[:DATA]?
Query upper limit
:SOURce <NRf>
Set digital output (0 to 15)
0
:SOURce?
Query source (digital output value)
:LOWer
Path to configure lower limit:
[:DATA] <n>
Specify lower limit (-9.999999e35 to
-1
+9.999999e35)