Section 7: Triggering
Model 6517B Electrometer Reference Manual
7-20
6517B-901-01 Rev. C / August 2015
The trigger link connections for this test system are shown in the following figure. The trigger link of
the Model 6517B is connected to trigger link of the Model 7001/7002 Switch System. Notice that only
one trigger link cable is needed. For this example, the Models 6517B and 7001/7002 are configured
as follows:
Figure 83: Trigger link connections (asynchronous example)
Model 7001 or 7002:
Idle State:
Reset = :INIT:CONT OFF*
Scan List = 1!1-1!10,
Arm layer:
Arm spacing = Immediate*
Arm count = 1*
Arm trigger control = Acceptor*
Scan layer:
Scan spacing = Immediate*
Number of scans = 1
Scan trigger control = Acceptor*
Channel Layer:
Channel spacing = TrigLink
Trigger link mode = Asynchronous
Input line = #4
Output line = #3
Number of channels = 20
Channel trigger control = source*
* Indicates that the setting is the RESET (and factory) default condition.
Содержание 6517B
Страница 136: ...Section 6 Test sequences Model 6517B Electrometer Reference Manual 6 8 6517B 901 01 Rev C August 2015...
Страница 180: ...Section 9 Filters and math Model 6517B Electrometer Reference Manual 9 4 6517B 901 01 Rev C August 2015...
Страница 492: ...In this appendix Information E 2 Appendix E IEEE 488 conformance...