background image

Safety

Introduction

Model 4200A-SCS Parameter Analyzer

Quick Start Guide

 

              

 

4200A-SCS

Summary of Contents for 4200A-SCS

Page 1: ...Safety Introduction Model 4200A SCS Parameter Analyzer Quick Start Guide 4200A SCS...

Page 2: ...require protection for high transient overvoltages often associated with localAC mains connections Certain Keithley measuring instruments may be connected to mains These instruments will be marked as...

Page 3: ...ecomes contaminated and operation is affected the board should be returned to the factory for proper cleaning servicing Safety precaution revision as of March 2016 Do not touch any object that could p...

Page 4: ...80 relative humidity non condensing Storage temperature 15 C to 60 C 5 F to 140 F 5 to 90 relative humidity non condensing Carefully consider and configure the appropriate output off state source lev...

Page 5: ...4200A SCS Parameter Analyzer performs lab grade DC and pulse device characterization real time plotting and analysis with high precision and sub femtoamp resolution The 4200A SCS offers the most advan...

Page 6: ...dware 4 Remove the packaging insert 5 Remove the 4200A SCS 6 Inspect the instrument for any obvious signs of physical damage Report any damage to the shipping agent immediately The 4200A SCS is heavy...

Page 7: ...ntegrated mouse Clarius software USB flash drive Power line cord 236 ILC 3 safety interlock cable 4200A SCS Quick Start Guide not shown this document Refer to the packing list for additional items tha...

Page 8: ...instrument Important test system safety information This system contains instruments that can produce hazardous voltages It is the responsibility of the test system installer maintenance personnel an...

Page 9: ...the instruments in your system If you need voltages greater than 40 V you must also connect the exterior of the test fixture enclosure to protective earth safety ground Take care to ensure that the w...

Page 10: ...interlock circuit is closed asserted and SMU 200 V ranges are enabled Conversely the interlock circuit is open de asserted when the lid is open and the SMU 200 V ranges are disabled High voltage test...

Page 11: ...onnect the other end of the connector to the test fixture Test connections If you are testing discrete devices you need a test fixture that is equipped with 3 lug triaxial connectors The 8101 PIV test...

Page 12: ...tlets When proper connections are made the instrument chassis is connected to power line ground through the ground wire in the power cord In addition a redundant protective earth connection is provide...

Page 13: ...d of the line cord into a properly grounded AC line power receptacle 3 Turn on the 4200A SCS by pushing the power switch to the lit position The instrument starts up When first starting a Clarius appl...

Page 14: ...ot properly reject powerline measurement noise To change the powerline frequency 1 Close Clarius 2 Run KCon 3 Select KI 4200A SCS in the Configuration Navigator The left pane is the Configuration Navi...

Page 15: ...Safety Introduction FAQs Next steps Connect...

Page 16: ...the mini triaxial cable 4200A MTRX from the SMU What should I do The mini triaxial connector is a locking connector To remove it pull the knurled part of the connector back A configuration error has...

Page 17: ...t will help you with the instrument 4200A SCS User Manual Contains basic information about using the system plus application based examples that will help familiarize you with the instrument 4200A SCS...

Page 18: ...s 1 800 1601 0077 Poland 41 52 675 3777 Portugal 80 08 12370 Republic of Korea 82 2 6917 5000 Russia CIS 7 495 6647564 Singapore 800 6011 473 South Africa 41 52 675 3777 Spain 00800 2255 4835 Sweden 0...

Reviews: