20
Table Display
If the numeric data is displayed on the home display, the data every minute, at
NIBP measurement, and at alarm occurrence will be stored and displayed as
table data.
Display interval, condition, and order can be changed on the “Table Setup”
window.
Place the jog dial marker on the
(Table/Trend) icon and press the jog dial to
display the table window.
The waveform on the waveform 1 display area will always be displayed.
1
2
3
4
5
6
7
8
1 Table Data Printing Date
Displays the printing date of the oldest data on
the table.
2 Display Selection
Displays the current display condition.
When selected, the display switches in the order
shown below.
Standard/NIBP/Alarm/St NIBP/
St Alarm/ All
3 Display
Interval
Displays
the current display interval.
When selected, the display interval switches in
the order shown below.
1 minute/2 minutes/5 minutes/10 minutes/15
minutes/30 minutes/60 minutes
4 Data at Alarm Occurrence
The data at alarm occurrence will be displayed
with a red background color.
↑
: upper limit has been exceeded
↓
: lower limit has been exceeded
5 Previous Screen
Displays previous table data.
6 Table Bar
The position and proportion of the stored data
are displayed through the position and width of
the slider.
7 Next Screen
Displays next table data.
8 Newest Screen Display
Displays the newest table data.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......