2
-
30
●
Parameter Alarm
For each parameter, ON/OFF of alarm and upper/lower alarm limit can be set.
When the set limit is exceeded, an alarm will generate.
WARNING
Set the appropriate upper and lower alarm limit for each
parameter according to the monitoring condition.
When the upper/lower alarm limit of the individual parameter
is set to “---“, the alarm will not be generated even if the
individual parameter alarm is set to ON. Be cautious when
setting them OFF.
CAUTION
If the measurement data of alarm generating parameter returns
within the normal range, alarm sound will automatically cease,
and alarm state will be cancelled.
●
Alarm Behavior Setup
Set the following items related to the alarm operation in advance.
・
Alarm Silence Time
(1 min. / 2 min.)
・
Alarm Sound Suspend
(60 min. / 90 min. /120 min.)
・
Alarm Sound Suspend Reminder
(OFF/ 3 min./ 10 min.)
・
Alarm Indicator
・
HR Zero Alarm (ON/ OFF)
Reference
For setup procedure, refer to “8. Maintenance/Maintenance Menu”.