Filter Mode Setup
The waveform frequency characteristic can be selected from “Monitor Mode”,
“ESIS Mode”, or “Diagnosis Mode” according to the monitoring purpose.
(1) Diagnosis Mode (Frequency Characteristics: 0.05-40Hz)
As the low frequency is set to 0.05Hz, select this mode when monitoring an ECG
with low frequency component.
(2) Monitor Mode (Frequency Characteristics: 0.5-40Hz)
This is the standard mode for ECG monitoring. The upper frequency is set to
40Hz to reduce artifacts caused by EMG, etc.
(3) ESIS Mode (Frequency Characteristics: 1.6-15Hz)
The upper frequency is limited to 15Hz, so that it can largely reduce
high-frequency artifacts.
1 Press the
(Menu) key, and select “Parameter Setup” >“HR/PR
Setup”>”Filter Mode”.
The “Filter Mode” window will be
displayed.
2 Select from [Diagnosis Mode] / [Monitor Mode] / [ESIS Mode].