xxii
CAUTION
Check the condition of cuff-applied part on the patient
during measurement so that the blood circulation will not
be blocked over long period of time by the squashed or
bent cuff hose.
Check the patient's condition constantly while measuring
over a long period of time with interval of 2.5 minutes or
less. Also, periodically check the blood circulation while
performing periodic measurement over a long period of
time. Congestion or rash may occur at the measuring site.
Make sure to check the patient's condition constantly when
repeatedly using continuous measurement as it may
cause dysfunction of patient's circulation.
When the cuff is not applied to the patient, pay attention
not to leave the cuff unattended. If periodic or continuous
measurement is set, the cuff will automatically inflate and
may cause the rubber bag inside the cuff to burst. When
not performing the NIBP measurement, set the NIBP
measurement interval OFF and disconnect the air hose
from the NIBP connector.
The following factors may affect the NIBP value.
- Body motion, arrhythmia, convulsion
- Continuous noise such as cardiac massage
- Periodic electromagnetic noise
For the following situation, measurements will be
terminated.
- When the measurement time has exceeded 160
seconds for adult and child, 80 seconds for neonate.
- When the inflation value has exceeded 300mmHg for
adult, 210mmHg for child, and 150mmHg for neonate.
If used with the incorrect patient classification, it will not
only cause erroneous measurement, but the inflating level
for the adult may be applied to child or neonate causing
dangerous situation to the patient.
When [1] minute is selected for “NIBP Measurement
Interval Setup”, the 1-minute interval measurement will
start from the time the selection is made. The
measurement will automatically stop after 12 minutes
(maximum of 15 minutes when re-measured), and
2.5-minute interval measurement will be set.
Selecting [Cont.] will start the continuous measurement.
The measurement will automatically stop after 12 minutes
(maximum of 15 minutes when re-measured), and
2.5-minute interval measurement will be set.
Alarm
If the measurement data of alarm generating parameter
returns within the normal range, alarm sound will
automatically cease, and alarm state will be cancelled.
The alarms are classified into Level 1 (High Priority), Level
2 (Medium Priority), Level 3 (Low Priority), Level 4
(Notification), and the alarm message will be displayed
according to the priority of level 1>level 2>level 3>level 4.
If multiple alarms are generated, the alarm message
display is switched every 2 seconds.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......