Alarm Occurrence
●
Alarm Occurrence
When the alarm occurs, it will be notified by a message display and an alarm
indicator.
If more than one alarm is generated, alarm messages will be alternately
displayed in 2-second intervals.
When a vital alarm occurs, the value inside the numeric data box will alternately
reverse between red and black.
When a vital alarm or technical alarm occurs, a message icon will be displayed
inside the numeric data box.
●
To Silence the Alarm
If the
(Alarm Silence) key is pressed at alarm occurrence, the alarm sound
can be silenced for fixed amount of time. This function will not affect the alarm
message.
If the alarm cause still remains at completion of silence time, the alarm sound will
generate again.
Also, if another alarm with the same or higher level occurs during the alarm
silence time, the alarm sound for the new alarm will occur.
During the alarm silence state, the
(Alarm Silence) icon will be displayed.