xxiii
CAUTION
Whether to use the SpO
2
SEC alarm function and its
threshold selection should be based on the patient's
clinical indication portent and medical evaluation.
If the SpO
2
alarm and SEC alarm setup is set to OFF, the
SEC alarm integral value will be set to 0.
System Configuration
Note that moving the jog dial marker on the Setup display
does not select any setup item. To select an item, press
the jog dial.
If the time/date is not correctly set, or changed during
monitoring, malfunction may occur with NIBP
measurement, periodic printing, trend and list data.
Pay attention not to set the alarm volume too low to avoid
missing any important alarms.
A password is required for changing the important
settings. For the password, refer to “Chapter 8
Maintenance”.
Maintenance
Never use strong-acidic cleaning solution to clean the
display unit. Do not wipe the display unit with a cloth or
gauze with coarse texture. Wipe the display with the soft
cleaning cloth provided as optional accessory or with an
eyeglass cleaning cloth.
Clean the equipment frequently so stains can be removed
easily.
To prevent injury, it is recommended to wear gloves when
cleaning the equipment.
Do not allow liquids or cleaning solution to enter the
equipment or connectors.
Do not use organic solvents, thinner, toluene and benzene
to avoid damaging the resin case.
Do not polish the housing with abrasive or chemical
cleaner.
When sterilizing the entire room using a spray solution,
pay close attention not to have liquids get into the
equipment or connectors.
Use only neutral detergent to clean the housing. Do not
use chemical cloth, scrub brush, abrasive, polishing
powder, hot water, volatile solvent and chemicals
(cleanser, thinner, toluene, benzine, benzol, and synthetic
detergent for house and furniture), or sharp-edged tools.
The surface resin coating may be damaged, resulting in
discoloration, scratches, and other problems.
Do not use autoclave sterilization.
Do not open the housing.
Avoid alcohol or other liquids from getting into the
equipment.
Replace the periodic replacement parts periodically as
specified.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......