Sight Inflation
When “Sight Inflation” is set to ON, the inflation target level will be automatically
estimated during the inflation, and starts to deflate after the target level is
reached. The inflation speed is slow but allows to detect any sudden increase of
blood pressure to prevent re-inflation and discomfort for the patient.
When “Sight Inflation” is set to OFF, it will inflate to the target level set according
to the previous measurement result.
NOTE
The sight inflation function can be used only during the NIBP
auto mode measurement.
The sight inflation function cannot be used when the patient
classification is "Neonate".
The sight inflation function cannot be used when performing the
1-minute interval measurement or continuous measurement.
When performing manual measurement/measurement at alarm
occurrence, it will inflate to the fixed value (Adult: 180mmHg,
Child: 140mmHg, Neonate: 110mmHg) regardless of the sight
inflation setting.
1 Press the
(Menu) key, and select “Parameter Setup” > “NIBP Setup”
> “Sight Inflation”.
2 Select [ON] or [OFF].
[ON] will turn ON the sight inflation function.
[OFF] will turn OFF the sight inflation function.
Pump Operation Mode
Select normal or silent operation for pump inflation operation.
1 Press the
(Menu) key, and select “Parameter Setup” > “NIBP Setup >
“Pump Operation Mode”.
2 Select [Normal] or [Silent].
[Silent] will inflate with decreased speed to reduce the pump inflating sound.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......