WARNING
The purpose of the RR zero alarm is to alert the user to
evaluate for the possible occurrence of apnea events by
identifying the absence of respiration. It is not intended to be
classified as an “Apnea Monitor” and will not identify the
condition creating the possible event. (Central, Obstructive or
Mixed)
SpO
2
Monitoring
When measuring the SpO
2
of patient with high fever or
peripheral circulatory insufficiency, check the sensor
attachment periodically and change the attachment site.
The temperature of attachment site will rise 2 to 3
C due
to the sensor heat which may result in compression
necrosis and burn injury.
For the following case, accurate measurement may not be
possible.
- Patient with excessive abnormal hemoglobin (HbCO,
MetHb)
- Patient with the pigment injected to the blood
- Patient receiving CPR treatment
- When a sensor is applied to a limb with NIBP cuff,
arterial catheter, or intracatheter
- When measuring at site with venous pulse
- Patient with body motion
- Patient with small pulse
NIBP Monitoring
Before the NIBP measurement, make sure the patient
classification ([Adult]/[Child]/[Neonate]) is properly
selected. Otherwise, correct measurement cannot be
performed, and congestion or other injury may result.
Night Mode Setup
When selecting [Silence] for Volume or [Time Only] for
Screen Brightness, pay attention not to miss any important
alarm by simultaneously monitoring the bed on other
monitors such as central monitor.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......