Precautions about Connections to Peripheral Devices
In the interest of safe and sufficient performance of this equipment, the
connection of other manufacturers’ equipment to the monitor is not authorized,
unless the connection is explicitly approved by Fukuda Denshi. It is the user’s
responsibility to contact Fukuda Denshi to determine the compatibility and
warranty status of any connection made to another manufacturer’s equipment.
WARNING
Regarding the connector with
mark, only the peripheral
devices specified by Fukuda Denshi should be connected with
the given procedure. Use of an unspecified device may cause
electric shock to the patient and/or operator due to excessive
leakage current.
CAUTION
All the peripheral device connectors on the DSL-8001 system
are isolated from the power supply, but the peripheral devices
are not isolated. To prevent danger of electric shock, always
position the peripheral devices away from the patient.
When connecting peripheral devices to the DSL-8001 system, it is the user’s
responsibility to verify that the overall system complies with EN 60601-1-1/IEC
60601-1-1, “Collateral Standard: Safety Requirements for Medical Electrical
Systems”.
Accessories and Optional Accessories
WARNING
Use only the cables specified by Fukuda Denshi.
Use of other cables may result in an increase in emission
(generation of unwanted electromagnetic energy from this
equipment) or decrease in immunity (resistance to
electromagnetic interference generated from other devices).
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......