xx
CAUTION
When the discharge procedure is performed, patient data
such as trend data will be initialized. The parameter and
alarm will be reset according to the setting on [Main
Menu]>[Maintenance]>[Backup Setup]>[Backup at
Discharge].
ECG Monitoring
This equipment does not have an arrhythmia analysis
function.
For a patient with arrhythmia, the heart rate may not be
counted accurately.
If different types of electrodes are used at the same time,
the difference between the polarization potential from each
electrode may interfere monitoring. Make sure to use
electrodes of the same type.
Do not reuse/re-sterilize the disposable electrodes.
ECG measurement part is Type CF applied part, but it is
not intended to directly apply on patient's heart.
The indication for continuous use of the electrode is about
one day.
Replace the electrode if the skin contact gets loosen due
to perspiring, etc.
When an electrode is attached to the same location for a
long period, some patients may develop skin irritation.
Check the patient's skin condition periodically and change
the electrode site as required.
Use only the specified relay cables, lead cables, and
electrodes.
The conductive parts of electrodes and associated
connectors for applied parts, including the neutral
electrode, should not contact other conductive parts
including earth.
Verify proper electrode placement, lead, waveform size,
and filter mode selection. If not properly selected, it may
cause erroneous detection.
Respiration Monitoring
When a defibrillator is used during respiration monitoring,
a large offset voltage will be placed on the ECG
electrodes, which may cause interruption of monitoring for
a few seconds.
SpO
2
Monitoring
If the nail is rough, dirty, or manicured, accurate
measurement will not be possible. Change the finger or
clean the nail before attaching the sensor.
If irritation such as skin reddening or skin fit appears with
the sensor use, change the attachment site or stop using
the sensor.
When fixating the sensor with tape, do not wind the tape
too strong. At the same time, check the blood flow
constantly so that congestion is not generated at the
peripheral.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......