8
-
24
RESP waveform is in the baseline position.
Cause
: Electrode is detached.
Solution : Reattach the electrode. If the electrode contact is poor, replace the
electrode.
The <RESP No Signal Detect> message is displayed.
Cause 1 : The amplitude of the respiration waveform is too low.
It is below 0.2 ohm which is outside the detection range.
Solution : Change the electrode site and increase the ECG amplitude.
Cause 2 : The electrosurgery-proof type ECG relay cable is used.
Solution : The impedance respiration can not be measured if the
electrosurgery-proof type ECG relay cable is used. If not using
during electrosurgery, use the standard ECG relay cable.
The <RESP Out of Range> message is displayed.
Cause 1 : The respiration rate is outside the measurement range.
Cause 2 : The electrode contact impedance is high.
Solution : Check if the electrodes are properly attached.
Reattach the electrode. If the electrode contact is poor, replace the
electrode.
The <RESP Suspended> message is displayed.
Cause : RESP measurement is suspended.
Solution : Press the (Menu) key, and select “Parameter Setup” > “RESP Setup”
> “RESP Measurement”, and select [ON].
The measurement data is displayed as “xxx”.
Cause
: The respiration rate is outside the measurement range.
Solution : Check if the electrodes are properly attached.
Reattach the electrode. If the electrode contact is poor, replace the
electrode.
The <RESP Communication Error> message is displayed.
Cause : Communication error with the ECG board has occurred.
Solution : Reboot the equipment. If the same message is repeatedly displayed,
a failure of the equipment can be considered. Cease the
measurement and contact our service representative.
Non-Invasive Blood Pressure (NIBP)
The cuff is not inflated although the pump is operating.
Cause 1 : The air hose is not firmly connected, and air is leaking.
Solution : Check if the air hose is properly connected.
Cause 2 : The cuff size does not correspond with the selected patient
classification.
Solution : Check if the cuff size corresponds with the selected patient
classification.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......