9
-
19
Item
Default
Backup
Adult
Child
Neonate
Pressure Unit
mmHg mmHg mmHg
○
Notification of Measurement
OFF
OFF
OFF
○
NIBP Data Display Time
120 min.
120 min.
120 min.
○
NIBP
Alarm
ON ON ON
○/△
Measurement Interval Setup
OFF
OFF
OFF
○/△
TEMP
TEMP Unit
℃
℃
℃
○
TEMP
Alarm
ON ON ON
○/△
Alarm
HR/PR Alarm Upper Limit
120 bpm
160 bpm
200 bpm
○/△
HR/PR Alarm Lower Limit
50 bpm
75 bpm
100 bpm
○/△
RESP Alarm Upper Limit
OFF
OFF
OFF
○/△
RESP Alarm Lower Limit
8 Bpm
8 Bpm
30 Bpm
○/△
TEMP Alarm Upper Limit
OFF
OFF
OFF
○/△
TEMP Alarm Lower Limit
OFF
OFF
OFF
○/△
SpO
2
Alarm Upper Limit
OFF
OFF
OFF
○/△
SpO
2
Alarm Lower Limit
90 %
90 %
90 %
○/△
NIBP SYS Alarm Upper Limit
160 mmHg
120 mmHg
90 mmHg
○/△
NIBP SYS Alarm Lower Limit
90 mmHg
70 mmHg
40 mmHg
○/△
NIBP MAP Alarm Upper Limit
OFF
OFF
OFF
○/△
NIBP MAP Alarm Lower Limit
OFF
OFF
OFF
○/△
NIBP DIA Alarm Upper Limit
OFF
OFF
OFF
○/△
NIBP DIA Alarm Lower Limit
OFF
OFF
OFF
○/△