3
-
9
To Measure the SpO
2
This section describes usable types of SpO
2
probes and how to attach the
probes.
WARNING
When measuring the SpO
2
of a patient with high fever or
peripheral circulatory insufficiency, check the sensor
attachment periodically and change the attachment site. The
temperature of the attachment site will rise 2 to 3
C due to the
sensor heat which may result in compression necrosis and
burn injury.
For the following case, accurate measurement may not be
possible.
Patient with excessive abnormal hemoglobin (HbCO,
MetHb)
Patient with pigment injected to the blood
Patient receiving CPR treatment
When a sensor is applied to a limb with NIBP cuff, arterial
catheter, or intracatheter
When measuring at site with venous pulse
Patient with body motion
Patient with small pulse
Do not connect a sensor or cable which is not authorized by
Fukuda Denshi to any I/O connector. If done so by mistake, the
equipment cannot deliver its maximum performance and the
connected units may be damaged, resulting in a safety hazard.
CAUTION
If irritation such as skin reddening appears with the sensor use,
change the attachment site or stop using the sensor.
Even a short duration of attachment may inhibit the blood flow
and generate compression necrosis and burn injury.
Check the sensor attachment site constantly every 4 hours
when probes or reusable sensor are used, and at least every 8
hours when a single-patient-use sensor is used. Be especially
careful of a patient with bad perfusion. If the sensor attachment
position is not changed constantly, skin irritation or skin
necrosis due to compression may be developed. For a patient
with bad perfusion, check the sensor attachment position at
least every 2 hours.
As skin for neonate/ low birth weight infant is immature, change
the sensor attachment site more frequently depending on the
condition.
Direct sunlight to the sensor area can cause measurements
error. Place a black or dark cloth over the sensor.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......