9
-
17
Item
Setup Value
Note
Main Menu/Maintenance/Alarm Behavior Setup
Alarm Silence Time
1 min. / 2 min.
Alarm silence duration can be
selected.
Alarm Sound Suspend
60 min. / 90 min. /120 min.
Alarm sound suspend duration
can be selected.
Alarm Sound Suspend
Reminder
OFF/ 3 min./ 10 min.
Interval to generate a warning
tone to remind the alarm sound
suspend condition can be
selected.
Alarm Indicator
Alarm Level 1-3 / Alarm Level
1-2 / Alarm Level 1 / OFF
Alarm level to activate the alarm
indicator can be selected.
HR Zero Alarm
OFF / ON
ON/OFF of HR zero alarm
function can be selected.
Main Menu/Maintenance/Backup Setup
Check Discharge at
Power ON
OFF / ON
Whether or not to display the
confirmation window for the
discharge operation can be
selected.
Automatic Discharge
Time
OFF / 10 min. / 1 hour / 2 hours /
4 hours
Power OFF duration to
automatically perform the
discharge operation can be
selected.
Backup at Discharge
OFF / ON
Whether or not to store the setup
data when a patient is
discharged can be selected.
Monitor Suspend at
Discharge
OFF / ON
Whether or not to suspend
monitoring when a patient is
discharged can be selected.
Measurement Interval at
Discharge
Off / 2 min. / 2.5 min. / 5 min. / 10
min. / 15 min. / 30 min. / 60 min. /
120 min.
NIBP measurement interval at
discharge can be selected.
Main Menu/Maintenance/Night Mode Setup
Night Mode
Manual / Auto
Whether to start the night mode
manually or automatically can be
selected.
Automatic Start Time
00:00 to 23:59
The time to automatically start
the night mode can be set.
Automatic End Time
00:00 to 23:59
The time to automatically end
the night mode can be set.
Volume
No Change / Quiet / Very Quiet /
Silence
Volume during the night mode
can be selected.
Screen Brightness
No Change / Slightly Dark / Dark
/ Time Only
Screen brightness during the
night mode can be selected.
Alarm Indicator
OFF / ON
ON/OFF of alarm indicator
operation during the night mode
can be set.
Setup Item (Level 4)
Item
Setup Value
Note
NIBP Measurement Interval Setup
NIBP Measurement
Interval
OFF/ Cont. / 1 min./ 2 min./ 2.5
min./5 min./ 10 min./ 15 min./ 30
min./ 60 min./120 min.
NIBP measurement interval can
be selected.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......