xxx
Compliance to the Electromagnetic Immunity (2)
The DSL-8001 system is intended for use in the electromagnetic environment
specified below. It is recommended that the DSL-8001 system be used in such
an environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Portable and mobile RF
communications equipment should
be used no closer to any part of the
DSL-8001, including cables, than
the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended Separation Distance
Conducted RF
IEC61000-4-6
3Vrms
150kHz to 80MHz
3Vrms
d = 1.2
P
Radiated RF
IEC61000-4-3
3V/m
80MHz to 2.5GHz
3V/m
d = 1.2
P
80MHz to 800MHz
d = 2.3
P
800MHz to 2.5GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
, should
be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Note 1 : At 80MHz and 800MHz, the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast can not be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered.
If the measured field strength in the location where the DSL-8001 system is used exceeds
the RF compliance level above, observe to verify whether the DSL-8001 system operates
normally in this environment. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the DSL-8001 system.
b
Over the frequency range 150kHz to 80MHz, field strength should be less than 3V/m.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......