Alarm at NIBP Measuring
When the SpO
2
sensor and the NIBP cuff are placed on the same limb for
measurement, an improper alarm may be generated because the cuff inflation
restricts the blood flow which disables the correct detection of the SpO
2
and PR
during the NIBP measurement.
1 Press the
(Menu) key, and select “Parameter Setup” > “SpO2 Setup”
> “Alarm at NIBP Measuring”.
2 Select [ON] or [OFF].
[ON] : SpO
2
alarm and PR alarm will be generated even during NIBP
measurement.
[OFF] : SpO
2
Alarm and pulse rate alarm will not be generated until the NIBP
measurement is complete.