RR Zero Alarm
Select whether or not to use the RR zero alarm function.
1 Press the
(Menu) key, and select “Parameter Setup” > “RESP Setup”
> “RR zero Alarm”.
2 Select [ON] or [OFF].
[ON] will generate an alarm when the RR value is zero. The threshold to generate
the RR zero alarm can be set on “RR zero Alarm Limit” which is explained in next
section.
WARNING
The purpose of the RR zero alarm is to alert the user to evaluate
for the possible occurrence of apnea events by identifying the
absence of respiration. It is not intended to be classified as an
“Apnea Monitor” and will not identify the condition creating the
possible event. (Central, Obstructive or Mixed)
RR Zero Alarm Limit
1 Press the
(Menu) key, and select “Parameter Setup” > “RESP Setup”
> “RR zero Alarm Limit”.
The “RR zero Alarm Limit” window
will be displayed.
2 Set the threshold to generate the RR zero alarm.
Aligning the highlight and pressing the jog dial will change the numeric color to
yellow enabling selection of value by turning the jog dial.
The selectable range is from 5 seconds to 60 seconds and can be set in 1
second increment.