4
-
3
●
Pacemaker
If [ON] is selected for "Pacemaker", the monitor will detect the pacing pulse
(pacemaker pulse) to perform the following process.
・
The artificial pacemaker pulse will be displayed.
・
When pacing waveform does not appear (pacing failure), erroneously
detecting the pacemaker pulse as QRS will be prevented.
1 Press the
(Menu) key, and select “Patient Setup”.
2 Select [ON]/[OFF] for the “Pacemaker”.
Monitor Suspend Setup
When the monitoring is suspended, all measurement functions, and table and
trend data storage will temporarily stop.
While the monitoring is suspended, <Monitor Suspended> is displayed on the
screen.
1 Press the
(Menu) key, and select “Patient Setup” > “Monitor
Suspend”.
The monitor suspend confirmation
window will be displayed.
2 Select [Yes] to suspend monitoring.
<Monitor Suspended> message and
[Resume] will be displayed.
On the home display, numeric data
and waveform displays will be
suspended.
If [No] is selected, monitoring will not be suspended and the confirmation window
will close.
3 To resume monitoring, press the jog dial or select [Resume].
If NIBP measurement interval setting is [Cont.] or [1] minute before the
monitoring is suspended, it will change to [2.5] minute after the monitoring is
resumed.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......