Ventilator Connection
By connecting this equipment to a ventilator, ventilator alarm can be notified to
the central monitor through wireless network or TCON system.
This section describes the procedure to connect this equipment and ventilator,
and to input the ventilator alarm.
NOTE
Only one ventilator can be connected for each DSL-8001 system.
When connecting a ventilator, check the corresponded software version of the
ventilator.
Ventilator Corresponding
Software
Version
Servo-i
v1.5 / v2.0 / v3.0
Servo-s
v2.0 / v3.0
CAUTION
The ventilator operation should be performed by a well-trained
or authorized person.
When connecting this equipment and the ventilator, use only
the specified connection cable.
Verify that this equipment and the ventilator are properly
connected.
When connecting the cable, verify that the main power of this
equipment and the ventilator are OFF.
Ventilator Connection
1 Turn OFF the power of this equipment and ventilator.
2 Connect one end of the optional ventilator connection cable to the
ventilator and the other side to the serial connector on the left side of this
equipment.
Ventilator Connection
Cable
Servo-i / Servo-s
CJ-502
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......