The <SpO
2
No Pulse Detect> message is displayed.
Cause
: The amplitude of the pulse waveform is too low, or the sensor is not
positioned correctly.
Solution : Check if the light emitting and receiving parts of the sensor LED are
aligned.
The <SpO
2
Communication Error> message is displayed.
Cause
: Communication error with the SpO
2
module has occurred.
SpO
2
sensor was connected with the power turned ON.
Solution : Make sure to connect the SpO
2
sensor before turning ON the power.
If the same message is repeatedly displayed, a failure of the
equipment can be considered. Cease the measurement and contact
our service representative.
The <SpO
2
Motion Artifact> message is displayed.
Cause
: There is excessive body motion from the patient which disables
correct measurement.
Solution 1 : Have the patient lie still.
Solution 2 : Relocate the sensor, or change the sensor to which the body motion
will have less influence.
Temperature (TEMP)
The measurement data is displayed as “
----
”.
Cause
: The temperature measurement is outside the measurement range.
Solution : Check if the temperature probe is properly inserted/attached.
The <TEMP Disconnected> message is displayed.
Cause
: When the temperature sensor is disconnected during TEMP
monitoring, this message will be displayed.
Solution : Check if the temperature probe is properly inserted/attached.
The <TEMP Out of Range> message is displayed.
Cause
: The temperature measurement is outside the measurement range.
Solution : Check if the temperature probe is properly inserted.
The <TEMP Unknown Sensor> message is displayed.
Cause 1 : The 700 series temperature probe may be used.
Solution : Use the 400 series temperature probe for measurement.
Cause 2 : There is a contact failure of the temperature probe.
Solution : Check if the temperature probe is properly inserted.
The <TEMP Communication Error> message is displayed.
Cause : Communication
error
with
the TEMP board has occurred.
Solution : Reboot the equipment. If the same message is repeatedly displayed,
a failure of the equipment can be considered. Cease the
measurement and contact our service representative.
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......